Thermo Scientific Q Exactive Focus

Thermo Scientific Q Exactive Focus

  • Mass spectrometer with quadrupole precursor ion selection and a high-resolution accurate-mass Orbitrap mass analyser
  • Appropriate for LC-MS/MS applications (HCD cell)
  • Resolving power of up to 70,000 (FWHM) at 200 m/z
  • 50 to 3000 m/z
  • Sub ppm mass accuracy
  • Rapid polarity switching
  • Linear dynamic range up to six orders of magnitude
  • Acquisition possible via Selected Ion Monitoring (SIM), Parallel Reaction Monitoring (PRM), and Data-Independent Acquisition (DIA)

In Stock

Description

Thermo Q Exactive Focus Mass Spectrometer

Benchtop LC-MS/MS system with quadrupole for precursor ion selection, higher-energy collisional dissociation (HCD) cell and Orbitrap high resolution accurate mass analyser. This system is particularly appropriate for the routine analysis of small molecules, and can be used for both targeted and untargeted applications. We are happy to provide Xcalibur software, an LC system and/or rough pump as required. Warranty options are available, please get in touch!

Performance:

  • Resolving power 70,000 at 200 m/z
  • Mass range 50 to 3000 m/z
  • Scan rate up to 12 Hz at resolution setting of 17,500 at 200 m/z*
  • Mass accuracy Internal: <1 ppm RMS External: <3 ppm RMS
  • Dynamic range >5000:1
  • Linear Dynamic >1,000,000 range
  • Polarity switching One full cycle in <1 sec (one full positive mode scan and one full negative mode scan at a resolution setting of 35,000)
  • Analog inputs One (1) analog input (0–1 V) One (1) analog (0–10 V)

 Operation modes:

  • Full MS with high-resolution accurate- mass detection
  • Selected Ion Monitoring (SIM) with high-resolution accurate-mass detection
  • Parallel Reaction Monitoring (PRM) for improved screening and quantitative confidence
  • Data-Independent Acquisition (DIA) for unbiased identification and quantitative analysis
  • All-Ion Fragmentation (AIF) in the HCD collision cell with high-resolution accurate-mass detection
  • In-source fragmentation of all ions
  • Positive/negative ion switching on chromatographic timescale
  • Timed SIM for scheduled data acquisition of the targets of interest

Hardware Specifications:

Ion Source

  • Removable metal ion transfer capillary provides vent-free maintenance

Ion Optics

  • RF lens stacked-ring radio frequency ion guide captures and efficiently focuses the ions into a tight beam. Large variable spacing between electrodes allows for better pumping efficiency and improved ruggedness.
  • Bent flatapole ion guide reduces noise by preventing neutrals and high-velocity clusters from entering the quadrupole

Quadrupole Mass Filter

  • Thermo Scientific™ HyperQuad™ mass f ilter provides increased ion transmission
  • Variable precursor ion isolation width selection from 0.4 Da to full mass range HCD Cell C-Trap

Vacuum system

  • Differentially pumped vacuum system with final vacuum <1 × 10-9 mbar
  • Two split-flow turbomolecular pumps
  • Seven vacuum regions

Orbitrap Mass Analyser

  • Nitrogen-filled C-Trap
  • Highly efficient ion transfer to Thermo Scientific™ Orbitrap™ mass analyzer
  • Straight multipole collision cell for HCD
  • Orbitrap mass analyzer with 5 kV central electrode voltage
  • Low-noise image current preamplifier
  • 16-bit signal digitalization
  • Automatic gain control